Peristaltic Pumps - Lambda - Operating Instructions

Peristaltic Pumps — operating instructions navigator
                   1. Installation of the Peristaltic Pump
                   2. Recommendations
                   3. Lambda Peristaltic Pumps - For Your safety
                   4. Construction Advantages Of Lambda Pumps
                   5. Flow Rate Diagrams
                   6. Accessories and spare parts
                   7. Specifications
                   8. Guarantee
                  

The successful design and well-proved mechanics of the PRECIFLOW peristaltic pump has been extended for by flow rate programming (HiFLOW and MULTIFLOW).
Up to 99 steps of time and flow rate can be easing programmed allowing the creation of any flow rate profile. The maximum flow rate has been increased from 600ml/hr up to 3000 ml/hr (HiFLOW). Up to now it has not been possible to produce pumps with such a high flow rate in just 1 l of instrument volume.

1. Installation Of The Peristaltic Pump


Figure 1

Figure 1
  • Plug the connector of the power supply into the corresponding socket on the backside of the pump and secure it in place by rotation of the ring.

  • Plug the power supply into an AC mains outlet (90—240 V / 50—60 Hz). After a short beep signal the display will be illuminated. The last used settings will appear on the display.

  • Remove the clear PVC cover by rotating it to the left or to the right (figure 1).

  • Press the ON/OFF button and select the sense of rotation of the peristaltic pump (button <|>) so that the rotor will turn in a clockwise direction.


Figure 2

Figure 2
  • Set the speed control to about 500 (PRECIFLOW and MULTIFLOW) and 100 (HiFLOW) by pressing the corresponding buttons under the LED display.

  • Press the tubing into the back slot on the top of the peristaltic pump. Thin tubing should be pushed completely to the bottom of the slot. Guide the tubing around the outside of the slowly turning plastic bearing towards the front slot (figure 2). Then press the tubing into the front slot to secure it (figure 3).

  • Replace the transparent cover (figure 4) by placing it on the top of the peristaltic pump and turning it so that the steel ball embedded in the left-front corner fits into the corresponding notch in the PVC cover. (When the cover is gently pressed the bearings will automatically adjust themselves).


Figure 3

Figure 3

Figure 4

Figure 4


 ↑ Up

1.1 ON/OFF button

By pressing the ON/OFF button the peristaltic pump is switched on or off. The internal memory will show the last used speed and flow direction setting.

1.1.1 Setting the flow rate

The flow rate of liquid through the peristaltic pumps depends upon the internal diameter of the tubing and the pump speed.

All Lambda peristaltic pump have been constructed for tubing with an internal diameter ranging from 0.5 to 4 mm with a wall thickness of approximately 1 mm. The best results have been obtained with silicon tubing, but tubing made from other materials with similar elasticity can also be used.

With the speed control buttons under the LED display the speed of rotation can be selected. The volume of liquid pumped per hour for tubing of different diameter and different speed settings is shown in the flow rate diagram. The flow rate can be set between 0.01 and 600 for PRECIFLOW and MULTIFLOW. The HIFLOW and MAXIFLOW peristaltic pumps have a flow rate range of up to 3 000 ml/hour and 10 000 ml/hr respectively.

1.2 Setting of flow direction

The direction of pump rotation can be selected by the <|> button. The corresponding LED diode will be on.


 ↑ Up

1.3 Remote control

1.3.1 ON/OFF control


By interlinking the contacts no. 4 and 5 of the socket on the rear of the peristaltic pump, the pump will be blocked (both direction indicating LEDs will be switched off).

By interlinking the contacts no. 4 and 5 (see chapter 7 for the contact description) of the socket on the rear of the peristaltic pump (see figure 5), the pump will be blocked (both direction indicating LEDs will be switched off). The same effect will be obtained by applying a voltage of 3 to 12 V DC to the contact no. 5 (0 line must be connected to contact no. 3). (Sometimes a reversed logic for the remote control is required. Please contact us in this case).

1.3.2 Remote control of the pump speed

This peristaltic pump can be controlled over the whole range by an external signal (0—10 V, option 0—20 or 4—20 mA). The plus pole of the signal is connected to the contact no.1, 0 line to the contact no. 3. Press the button REMOTE on the front panel. The corresponding LED diode will go on and the display will indicate an approximate voltage of the external signal. This indication may become unstable when external connection is not made indicating the high sensitivity of the electronics.

Attention: For safety reasons the voltage of the external signal must not exceed 48 V to earth!


 ↑ Up

1.4 How to program the peristaltic pump (only HiFLOW and MULTIFLOW)

Up to 99 pairs of time and flow rate may be programmed in a simple way. The access to the program is opened by a simultaneous pressing of buttons REMOTE and RUN until the indication PGM will appear on display and both directions diodes (<|> button) will be illuminated.

(If you repeat this simultaneous pressing, the memory will be cleared and the indication cLE will appear on the display. For programming press both buttons again until PGM appears again).


  • Press the button ON/OFF. The indication F01 will appear for one second on the display indicating that you can select the first flow rate value. You can set the desired flow rate value (0 to 999 representing 0 to 100% speed of rotation) by pressing the buttons under the display. Press the ON/OFF button again. The indication t01 will appear for a second on the display indicating that you can program the time period of the first step in minutes. Select the desired time period for the first program step.

  • Press the button ON/OFF again. The indication F02 will briefly appear on the display. You can enter the desired flow rate. After this press ON/OFF button again. The indication t02 will briefly appear on the display. You can now set the time of the second step.

  • In a similar way up to 99 steps can be programmed.

  • After having entered the time of the last step press simultaneously buttons REMOTE and RUN until an indication End will appear on the display (one of the direction diodes will go off). It is not possible to end the program after programming the flow rate.


To start the program press button RUN.

To stop the running program press button RUN.

It is possible to stop the rotation of the peristaltic pump, to change the direction and speed of rotation during a running program step. This facilitates urgent changement of the tubing or allows reacting to another emergency situation. Do not forget to restore the right direction of flow rate and to switch the peristaltic pump on after you have finished your intervention. The time basis in the microprocessor is not stopped so that the total time of the step and of the program will not be affected by this modification. After the end of the running step the pump will automatically go on to the following step. The program is not modified by this emergency intervention.

After the last step of the program the peristaltic pump will continue the pumping with the flow rate of the last step until it is switched off. If you wish that the pump stops after the last step, program an additional step with flow rate 000 during for example one minute.

You can view the program by proceeding in the same way as during programming but without modifying of it.


 ↑ Up

2. Recommendations


  • It is advisable to use tubing with a small diameter and high-speed control setting for small flow rates rather than the opposite. This allows a finer selection of flow rates.

  • If possible use the clockwise rotation of the peristaltic pump. This results in lower friction and the pressure of the liquid is also lower (approx. 0.1 MPa). If a higher pressure is required (up to 0.15 MPa) use the opposite direction.

  • Periodically smear a small amount of petroleum jelly or similar grease on the inner side of the transparent PVC cover of the peristaltic pump. This will increase the reliability and the lifetime of your pump. Do not grease the tubing holding slots.

  • If as a consequence of tube breakage or some other accident liquid spills into the top of the peristaltic pump disconnect the pump from the mains and clean it by removing the liquid and rinsing it with water. The rotor can be completely removed by screwing a nut (size M4) to the axis of the rotor and pulling the rotor out by hand or with a pair of pliers. After cleaning, grease the axis and replace the rotor by pressing and rotating until the rotor engages on the motor axis.

  • Should you have any difficulties or questions concerning your peristaltic pump please contact our service office.


 ↑ Up

3. LAMBDA Peristaltic Pumps — For Your Safety


  • Thanks to use of plug in power supply giving only low voltage of 12 V DC the danger of electrical shock during the use of peristaltic pump has been virtually eliminated. This holds even for the case when an electro-conducting solution penetrates the pump.

  • The peristaltic pump is usually used in a vertical position. The pumps can also be stacked on each other allowing optimal use of your expensive laboratory areas.

  • If the peristaltic pump is not used for an extended period of time, disconnect it from mains. We use modern miniaturised switching power supply, which has a negligible consumption of electrical current when the pump is not used.


 ↑ Up

4. Construction Advantages Of Lambda Pumps


  • Instead of small rollers, which are used by most peristaltic pump producers, we use bearings of a larger diameter with glass beads. This reduces pulsation, friction and mechanical strain on the tubing leading to better pump performance.

  • The bearings glide over the tubing so gently, that unlike other peristaltic pumps, it is not necessary to prevent movement of the tubing by a special fixation. In this manner the lifetime of the tubing is considerably increased.

  • The pressure on the tubing is delivered individually through an off-centre lever and spring of stainless steel. This assures that the minimal pressure is applied to the tubing, which still guarantees good functioning of the peristaltic pump without unnecessary deformation of the tubing.

  • The spring also reduces the liquid pressure to approximately 1 bar (depending of the tubing diameter). This is useful when for any reason the line is blocked.

  • The double-asymmetric pump head reduces pulsation.

  • A high quality motor together with integrated microprocessor electronics assure a high precision of flow rate with no inertia while turning the peristaltic pump on and off.

  • The pump head is made from hard, chemically stable epoxide.

  • The pump dimensions have been minimised. Therefore LAMBDA peristaltic pumps are considerably smaller than other products of similar performance. The peristaltic pump is easy to use and saves space in the laboratory.

  • Remote control and possibility of flow integration increases the scope of use of the peristaltic pump in automatically controlled systems (e.g. fermentation, chemical synthesis, fraction collection, etc.)

  • The microprocessor allows up to 99 program steps. This opens the possibility to establish almost any flow rate profile. This is particularly useful for adding medium to the fermenter to feed the culture growing also exponentially. The growth rate and resulting culture activity are considerably increased


 ↑ Up

5. Flow Rate Diagrams


5.1 PRECIFLOW, MULTIFLOW

Flow rate diagram of the PRECIFLOW or HIFLOW Peristaltic Pump

Flow rate diagram of the PreciFlow or HiFlow Peristaltic Pump

5.2 HiFLOW

Flow rate diagram of the HiFlow Peristaltic Pump

Flow rate diagram of the HiFlow Peristaltic Pump

5.4 MAXIFLOW




 ↑ Up

6. Accessories And Spare Parts


6.1 Pump-Flow Integrator (Cat. no. 4803)

The LAMBDA peristaltic pumps are the only pumps on the market, which allow a simple but precise integration of the amount of liquid, which has been transported, by the pump.

The electric impulses, which move the stepping motor, are registered and transformed into a direct voltage. The voltage can be measured or recorded by common recorders or voltmeters.

In processes where the peristaltic pump is controlled e.g. by a pH-stat during a fermentation to keep the pH of the medium constant, it is frequently important to know when and how much acid or base was added. This data yields important information about the process, its kinetics, time of completion, etc. Another use of the Integrator is for measurement of enzyme activities (esterases, amidases, lactamases and other enzymes). The Pump-Flow Integrator can be placed under the peristaltic pump, to which it is connected by a cable to the 5-pole socket on the rear of the pump (figure 4).

6.2 Remote control cable (Cat. no. 4810)

6.3 Spare parts

Cat. no.
 
4805
PE bearings
4806
Stainless spring
4807
Off-centre lever
4808
Rotor
4809
PVC cover

 ↑ Up

7. Specifications


Flow rates:
0.01 to 600ml/hr (PRECIFLOW and MULTIFLOW), 0,02 to 3000 ml/hr (HIFLOW), and 0,02 to 10000 ml/hr (MAXIFLOW) using tubing (silicon or similar) with internal diameter from 0.5 to 4 mm, wall thickness 1 mm
Dimensions:
10.5 (W) × 9.5 (H) × 10.5 (D) cm
Weight:
1 kg (PRECIFLOW and MULTIFLOW), 1.2 kg (HIFLOW and MAXIFLOW)
Power source:
12 V / 8 W (PRECIFLOW and MULTIFLOW) and 12/ 24W (HIFLOW) using plug in power supply for 90—240 V / 50—60 Hz
Safety:
CE, meets IEC 1010/1 norm for laboratory instruments
Operating temperature:
0—40 °C
Operating humidity:
0—90 %
Remote control:
0—10 V, (option 0—20 or 4—20 mA)
! For safety reasons the voltage of the remote signal to earth must not exceed 48 V DC

INPUTS/OUTPUTS:

Image of connector

Contact no.: and cable colour code
1 + input remote speed control 0–10 V (yellow)*
2 step signal from stepping motor (0 and 12V) (grey)
3 earth, 0 V (green)
4 12 V + (brown)
5 + input remote ON/OFF (white)* 0 V = ON, 3—12 V = OFF (this logic can be inversed on demand) *(zero line connected to the contact no. 3)

 ↑ Up

8. Guarantee

LAMBDA PRECIFLOW and MULTIFLOW peristaltic pumps have a five-year guarantee for all labour and components when the instrument has been used according to our operating instructions and the advice given above.

LAMBDA HIFLOW and MAXIFLOW peristaltic pumps have a two-year guarantee for all labour and components when the instrument has been used according to our operating instructions and the advice given above.

Conditions of guarantee:

  • The instrument must be returned with a complete description of the defect or problem.

  • The customer will send the instrument to our service office.

  • Damage or loss during transport of items to us will not be compensated by Lambda.

  • Failure to fulfill these requirements will disqualify the customer from compensation.