| PRECIFLOW pump | MULTIFLOW pump | HIFLOW pump | MAXIFLOW pump | MEGAFLOW pump |
---|
Flow rate range | 0.2 µl/min - 600 ml/h | 0.2 µl/min - 600 ml/h | 1 µl/min - 3,000 ml/h | 3 µl/min - 10,000 ml/h | 0.02 ml/min - 60,000 ml/h |
---|
Programming locally on the display | N/A | up to 99 steps of speed and time | up to 99 steps of speed and time | up to 99 steps of speed and time | up to 99 steps of speed and time |
---|
Time resolution | N/A | 0 to 999 minutes in 1 minute steps; 0 to 99.9 minutes in 0.1 minute steps | 0 to 999 minutes in 1 minute steps; 0 to 99.9 minutes in 0.1 minute steps | 0 to 999 minutes in 1 minute steps; 0 to 99.9 minutes in 0.1 minute steps | 0 to 999 minutes in 1 minute steps; 0 to 99.9 minutes in 0.1 minute steps |
---|
Accuracy | ± 1% | ± 1% | ± 1% | ± 1% | ± 1% |
---|
Reproducibility | ± 0.2 % (electronics) | ± 0.2 % (electronics) | ± 0.2 % (electronics) | ± 0.2 % (electronics) | ± 0.2 % (electronics) |
---|
Tubing | Silicone tubing or other materials having similar elasticity; inner tubing diameter from 0.5 to 4 mm and tubing wall thickness of approx. 1 mm | Silicone tubing or other materials having similar elasticity; inner tubing diameter from 0.5 to 4 mm and tubing wall thickness of approx. 1 mm | Silicone tubing or other materials having similar elasticity; inner tubing diameter from 0.5 to 4 mm and tubing wall thickness of approx. 1 mm | Silicone tubing or other materials having similar elasticity; inner tubing diameter from 0.5 to 4 mm and tubing wall thickness of approx. 1 mm | Silicone tubing or other materials having similar elasticity; inner tubing diameter from 1 to 8 mm and tubing wall thickness of approx. 2 mm |
---|
Maximum pressure | approx. 0.1 MPa in clockwise rotation; approx. 0.15 MPa in counter-clockwise rotation | approx. 0.1 MPa in clockwise rotation; approx. 0.15 MPa in counter-clockwise rotation | approx. 0.1 MPa in clockwise rotation; approx. 0.15 MPa in counter-clockwise rotation | approx. 0.1 MPa in clockwise rotation; approx. 0.15 MPa in counter-clockwise rotation | approx. 0.18 MPa in clockwise rotation; approx. 0.2 MPa in counter-clockwise rotation |
---|
Motor | microprocessor controlled stepping motor | microprocessor controlled stepping motor | microprocessor controlled brushless long life BLDC motor with neodymium magnets | microprocessor controlled brushless long life BLDC motor with neodymium magnets | microprocessor controlled brushless long life BLDC motor with neodymium magnets |
---|
Speed control range | 0 to 999 | 0 to 999 | 0 to 999 | 0 to 999 | 0 to 999 |
---|
Interface | RS-485 or RS-232 (optional) | RS-485 or RS-232 (optional) | RS-485 or RS-232 (optional) | RS-485 or RS-232 (optional) | RS-485 or RS-232 (optional) |
---|
Power supply | 90–240 V/50–60 Hz AC plug-in power supply with DC 12V/12W output; possible field operation on 12 V accumulator | 90–240 V/50–60 Hz AC plug-in power supply with DC 12V/12W output; possible field operation on 12 V accumulator | 90–240 V/50–60 Hz AC plug-in power supply with DC 12V/50W output; possible field operation on 12 V accumulator | 90–240 V/50–60 Hz AC plug-in power supply with DC 12V/50W output; possible field operation on 12 V accumulator | 90–240 V/50–60 Hz AC plug-in power supply with DC 12V/65W output; possible field operation on 12 V accumulator |
---|
Dimensions | 10.5 (W) x 9.5 (H) x 10.5 (D) cm | 10.5 (W) x 9.5 (H) x 10.5 (D) cm | 10.5 (W) x 9.5 (H) x 10.5 (D) cm | 10.5 (W) x 9.5 (H) x 10.5 (D) cm | 18 (W) x 13 (H) x 16 (D) cm |
---|
Weight | < 1 kg | < 1 kg | 1.2 kg | 1.2 kg | 2.5 kg |
---|
Non-volatile memory | storage of all settings | storage of all settings | storage of all settings | storage of all settings | storage of all settings |
---|
Remote control | 0-10 V; (option 0-20 or 4-20 mA); foot switch | 0-10 V; (option 0-20 or 4-20 mA); foot switch | 0-10 V; (option 0-20 or 4-20 mA); foot switch | 0-10 V; (option 0-20 or 4-20 mA); foot switch | 0-10 V; (option 0-20 or 4-20 mA); foot switch |
---|
Safety | CE, meets IEC 1010/1 norm for laboratory instrument | CE, meets IEC 1010/1 norm for laboratory instrument | CE, meets IEC 1010/1 norm for laboratory instrument | CE, meets IEC 1010/1 norm for laboratory instrument | CE, meets IEC 1010/1 norm for laboratory instrument |
---|
Operation temperature | 0-40 °C | 0-40 °C | 0-40 °C | 0-40 °C | 0-40 °C |
---|
Operation humidity | 0-90% RH, not condensing | 0-90% RH, not condensing | 0-90% RH, not condensing | 0-90% RH, not condensing | 0-90% RH, not condensing |
---|
Warranty | 5 years | 5 years | 2 years | 2 years | 2 years |
---|