Type: | LAMBDA VIT-FIT / VIT-FIT HP – microprocessor-controlled programmable syringe pump (infusion / withdrawal) |
---|
Programming: | up to 99 steps of speed and time |
---|
Time resolution: | 0 to 999 minutes in 1 minute steps or 0 to 99.9 minutes in 0.1 minute steps: time resolution can be selected individually for each program step |
---|
Accuracy: | ±1% |
---|
Reproducibility: | ± 0.2% (electronics) |
---|
Syringes: | glass, plastic, metal syringes from 5 μl to over 150 ml |
---|
Flow rate: | depends on the inner syringe diameter |
---|
Non-volatile memory: | storage of all settings |
---|
Maximum force: | VIT-FIT: 300 N (reducible by a switch to 80 N) VIT-FIT HP: 600 N (reducible by a switch to 160 N) |
---|
Motor: | microprocessor controlled brushless long life BLDC motor with neodymium magnets |
---|
Transmission: | efficient force transmission by a ball screw with highest mechanical load capacity of 12’000 N |
---|
Pusher travel: | 120 mm |
---|
Pusher travel rate:
| |
---|
Minimum: | 0.08 mm/min |
---|
Maximum: | 80 mm/min |
---|
Speed control range: | 0 to 999 |
---|
Interface: | RS-232 or RS-485 (optional) |
---|
Power supply: | 95–240 V/50–60 Hz AC plug-in power supply with DC 12V/24W output; possible field operation on 12 V accumulator |
---|
Dimensions: | 26.5 cm x 12.5 cm × 13 cm (W × D × H) |
---|
Weight: | 3,2 kg |
---|
Safety: | CE, meets IEC 1010/1 norm for laboratory instruments |
---|
Operation temperature: | 0-40 °C |
---|
Operation humidity: | 0-90% RH, not condensing |
---|
Remote control: | 0-10 V; (option 0-20 or 4-20 mA) |
---|