3.1 Principle of programming

Laboratory practice shows that programming of instruments equipped with microprocessors are not simple and mistakes are easily made, especially when such instruments are used occasionally.

We have developed a new method, which should eliminate such problems and make the programming easy for all types of tube racks.

Figure 3.1-1 Magnetic coding band used for the fraction
positioning.
Figure 3.1-2 The fraction positions are indicated by white lines
drawn on the black magnetic coding band as shown.
Figure 3.1-3 The beginning and the end of a row are indicated by short lines (about half the length of the normal stripes or less and ~5 mm broad) made on the lower edge (for the beginning) and upper edge (for the end) of the coding band, respectively.
Figure 3.1-4 The distance between the fraction rows is programmed
in a similar way. If you wish to switch off the collector automatically
after the last row, draw the last line broader (approx. 1 cm)

 

The white lines should be about 2 mm broad and full-length. These lines are detected by the photodetectors of the OMNICOLL fraction collector and the fraction collector stops at these positions. The actual fraction position is about 1 mm after the first edge of the white lines.

The beginning and end signals (at Figure 3.1-3) determine at the same time the position of the first and the last fraction in the row. The fraction position is about 2 mm after the first edge of the white line of the beginning and end stripes.

This simple coding principle allows the use of any tube racks and fraction stands or other recipients like bottles, 96-well plates, vials, etc. with the LAMBDA OMNICOLL fraction collector and sampler.

 

3.2 Ready to use Fraction collector with the supplied tube racks

The magnetic coding bands programmed with white lines for the desired tube racks are provided with the OMNICOLL Fraction collector.

Place the magnetic coding strip strictly horizontally into the corresponding X-axis slot!

Otherwise, the optical detectors could be damaged.

Figure 3.2-1 Place the longer and narrower coding band (Y-axis) on the left inner metal frame just under the gear-bar as shown. The first coding line should lie approximately 10cm from the front frame.
Figure 3.2-2 Pull the X-axis coding band holder from the
OMNICOLL control unit as shown.
.
Figure 3.2-3 Place the shorter and broader magnetic coding
band into the X-axis holder.
Figure 3.2-4 Insert the X-axis holder with coding band into the
desired slot of the control unit, with the white coding stripes facing up.

 

3.3 Programming the OMNICOLL for any tube racks or recipients

3.3.1 Coding X-axis for tube distance in row

Measure the distance between the tube centres of the desired rack and choose the number of fractions in a row.

The beginning and end signals at the lower or upper edge of the coding band informs the microprocessor whether the arm of the control unit is on right or left.
Figure 3.3-1 Place the coding bands on the tube rack and mark the fractioning position with a normal pencil.
.
Figure 3.3-2 Draw the white lines correctly on the pre-marked places. Reserve approximately 2 cm of the coding band for the beginning and end-position signals.
Figure 3.3-3 Mark the beginning and end signals on the lower and upper edge of the coding band respectively. Leave a minimum of about 6mm free space between the first and last line of the row and the beginning and end signals.
Figure 3.3-4 Place the coding band in the X-axis holder and insert into the desired slot of the control unit, with white stripes facing up.
.

 

3.3.2 Coding Y-axis for row distance

Measure the distance between rows. When several tube racks are used, pay attention to difference in distance between the last row of the first rack and the first row of the second rack.

Measure the distance between rows. When several tube racks are used, pay attention to difference in distance between the last row of the first rack and the first row of the second rack.

If the fraction collection should be stopped at the last row, then make the last line broader (approximately 1 cm thick). This functions as the stop signal.

Since the coding band is kept in place by magnetic force, it can be easily positioned.

The first row line of the Y-axis should be ca 10 cm from the front.

When the control unit is placed in the starting position on the front of the frame, the photoelectric detector on the left side of the control unit must be situated before the first line. Otherwise the control unit will stop at the position of the next row.
Figure 3.3-5 Mark the row distance with a pencil on the coding band for Y-axis and draw the white lines with the provided correction pen.
Figure 3.3-6 Place the coding band on the left internal side of the
metal frame just under the gear bar.

 

3.4 Control Panel of OMNICOLL Fraction Collector

 

3.4.1 Selection of linear, meander-like or row collection

The LAMBDA OMNICOLL fraction collector-sampler allows the following operation modes:

Line:

The fractions are collected from the left to the right. After the last fraction of the row, the collecting arm moves to the left to the first position of the next row.

Meander: The fractions are taken alternatively from the left to the right and in the following row from the right to the left and so forth.
Row:The collecting arm does not move within the row. The collector moves only from row to row. This collection mode is mainly used in the simultaneous (multi-channel) fraction collection mode.

In the row mode, the position of the fraction collecting arm can be adjusted by using the STEP button.

 A small difference in the fraction positions with respect to the tube centres may occur between fractions taken in opposite collection directions. If the fraction lines are not too broad (about 2 mm), this difference in position is of approx. 1 mm
Figure 3.4-1 Press the button LINE continuously while plugging the power supply into the mains.
.
Figure 3.4-2 “rouu” for row mode of collection appear on the display. Press the SELECTION (below COUNT / TIME) button to select the desired mode of collection.
Figure 3.4-3 “Mean” for meander-like mode of collection appear on the display.
Figure 3.4-4 “Line” for linear mode of collection appear on the display.
Figure 3.4-5 Confirm the selected mode of collection by pressing SET button.

 

3.4.2 Time resolution (0.1 or 1 minute)

The time resolution of the OMNICOLL fraction collector can be selected in the following way:

Figure 3.4-6 Press the button COUNT/TIME while connecting the power supply to the mains.
Figure 3.4-7 The actual time resolution will appear on the display “0.1M” or “1M”.
Figure 3.4-8 By pressing the button COUNT / TIME below the COUNT and TIME LEDs you can select to use 0.1 minute steps “0.1M” or 1 minute steps “1M”.
Figure 3.4-9 Confirm the choice by pressing the button SET.
.

 

3.4.3 Drops and count divider (Division co-efficient)

This function allows increasing considerably (by 60 times) the volume of fractions either using a drop counter (art. no. 6926) or motor impulsions counter (counts).

The divider can be set either to divide by one or divide by 60. In the first setting, every drop and every impulsion of motor are counted. In the second setting, sixty drops or impulsions give just one signal.

The volume of fractions can be also varied by the selection of tubing of internal diameters from 0.5 to 4 mm, when constant volume fractions are collected by counting pump impulsions (in “COUN” operation mode – COUNT LED is on, refer section 3.5)

Figure 3.4-10 Press the button START while plugging the power supply into the mains.
Figure 3.4-11 “div” and then “=01” or “=60” will appear on the display.
.
Figure 3.4-12 By pressing the COUNT/TIME button select the desired division factor.
Figure 3.4-13 Confirm the choice by pressing the button SET.

 

3.4.4 Address selection (for PC control)

When the OMNICOLL fraction collector and sampler has been equipped with the optional RS-232 interface, it can be controlled digitally, e.g. from a PC.

Figure 3.4-14 While keeping the direction button STEP pressed connect the fraction collector to the mains.
Figure 3.4-15 The message “A” and two numbers will appear on the display. This number from 00 to 99 is the current address of the fraction collector.
Figure 3.4-16 To change the address press the buttons Λ Λ Λ Λ under the display until the desired number is obtained.
Confirm the choice by pressing the button SET.
Figure 3.4-17 Confirm the choice by pressing the button SET.

 

3.5 Fraction collection according to the volume

If you use the OMNICOLL fraction collector together with LAMBDA peristaltic pumps PRECIFLOW, MULTIFLOW, HIFLOW, MAXIFLOW or MEGAFLOW, you can take fractions of precise volume (from 0.05 to 500 ml or 0.6 to 30 litres per fraction).

The heart of these pumps is a stepping motor or BLDC motor, controlled by a generator of electric impulses (microprocessor). After each impulse, the pump motor moves by one step. This movement displaces a very small and precise volume of liquid.

The collector counts these impulses and thus makes it possible to deliver an exact volume of liquid for each fraction. This method is an attractive alternative to the old drop counting procedure, where the volume is affected by the viscosity, surface tension etc. of the respective liquid.

Since the diameter of the tubing used in the peristaltic pumps affects the speed of the flow (flow rate), the pump has to be calibrated before use in order to establish the relation between the count number and the delivered liquid volume.

It is also possible to synchronize OMNICOLL fraction collector with third-party multi-channel or single channel peristaltic pumps (e.g.: Cole parmer, Ismatec, Flexicon, etc.). We can offer customized remote control cable and communication module for pump switching and RS-232 connection (Art. No.: 6911) for the desired peristaltic pump connection.

 

3.6 Calibration of the peristaltic pump and the OMNICOLL

The LAMBDA peristaltic pump is connected to the OMNICOLL fraction collector via the remote control cable (Art. No. 4810-s) and the communication module box (Art. No. 6911 or Art. No. 6929). The plug-in power supply is also connected to this connection box. (It is impossible to make a wrong connection because the connectors will not fit to the wrong sockets).

Figure 3.6-1 Connect the peristaltic pump remote control cable
to the communication module box.
Figure 3.6-2 With the COUNT/TIME button select the count modus
(the COUNT LED will be switched ON)
Figure 3.6-3 Select the desired count number with the button Λ Λ Λ Λ under the display, for example: 55.
Figure 3.6-4 Confirm the selection by pressing the SET button.
.
Figure 3.6-5 Select the speed control on the peristaltic pump
(e.g. 700), choose the pump rotation direction and switch the
pump ON.
Figure 3.6-6 Press the START button on the collector to start fraction collection. If the tubing was not filled completely with the liquid, for example water, take a second fraction.

Measure the volume of this fraction. This volume corresponds to the pre-set count number.

From this ratio (volume/count number) you can easily calculate the count number corresponding to the desired fraction volume.

Since tubing of 0.5 to 4 mm of internal diameter (with a tubing wall thickness ~1 mm) can be used in the LAMBDA peristaltic pumps, the fraction volume can be selected in a wide range. For example, when tubing with an internal diameter of 3 mm is used, then one count corresponds approximately to one drop.

 

3.7 Fraction collection with a time interval in-between fractions (“high”)

LAMBDA OMNICOLL fraction collector and sampler could be used for the collection of samples with a time interval between consecutive fractions varying between 0.1 min and 16.6 hours (999.9 min) or 1 min and 166 hours (9999 min).

This is particularly useful during fermentations and other biological and chemical processes. For this application, switch the microprocessor of the control unit to the “high” mode.

Selecting the high mode:

Figure 3.7-1 Press the SET button on the control unit and connect
the plug-in power supply to the mains.
Figure 3.7-2 Use the COUNT/TIME button to select the “high”
mode.
Figure 3.7-3 Confirm the selection by pressing the SET button
(long beep signal).

In the “high” mode you can successively select the parameters TIME, COUNT, PAUSE and NUMBER by pressing the COUNT/TIME button.

After selection of the parameter with the COUNT/TIME button, press the SET button.

After a short beep the name of the parameter will appear during one second, followed by last used value of the parameter.

You can change this value by means of the four buttons Λ Λ Λ Λ under the display. Press the button SET to confirm the new value (long beep).

Selecting the parameter COUNT
Figure 3.7 4 Selecting parameter COUNT
(volume).
[Count LED: ON;
Time LED: OFF;
Display: coUn]
Figure 3.7 5 Selecting the parameter TIME to
perform the program for a desired time.
[Count LED: OFF;
Time LED: ON;
Display: tiMe]
Figure 3.7 6 Selecting the parameter PAUSE
for introducing pauses in the program.
[Count LED: ON;
Time LED: ON;
Display: PAUS]
Figure 3.7 7 Selecting the parameter NUMBER
for collecting a number of samples.
[Count LED: OFF;
Time LED: OFF;
Display: nUMb]

After setting of the parameters, the fraction collection is started by pressing the START button.

The collector will take the number of samples as programmed under the NUMBER setting and will stop for the time programmed under the PAUSE setting. This cycle will go on until a STOP signal will be encountered. The LAMBDA peristaltic pump is activated automatically only during fraction collection.

When the fraction collection is controlled from the panel (i.e. not by remote control signal) the fraction collector will collect the number of fractions/samples introduced in the parameter NUMB of duration TIME or volume COUNT (if a drop counter or LAMBDA pumps are used), followed by the PAUSE, if a PAUSE was set.

Then, the fraction collector will repeat this till the STOP signal is encountered (i.e. STOP is pressed or the fraction collector is stopped by the end magnet). This collection mode allows for e.g. a line washing step before the relevant fraction is taken.

If operated by external (analog) remote control signal, the fraction collector will take the set number of fractions/samples of duration TIME or volume COUNT (if a drop counter or LAMBDA pumps are used), and will then stop till the reception of a new external remote control signal.

The parameter PAUSE is ignored in this operation mode.

Such a collection mode could be useful for taking of fractions/samples in a process where an alarm situation of a selected process parameter has occurred and the samples can be used for further analysis.

 

3.8 Multi-channel fraction collection – multiple stream sampling

LAMBDA OMNICOLL fraction collector and auto-sampler can be used for all kinds of chromatography such as normal or low pressure chromatography. It is also suitable for medium pressure chromatography sometimes called FPLC (Fast Protein Liquid Chromatography) or high pressure liquid chromatography (HPLC), also referred to as high performance liquid chromatography.

An interesting property of the LAMBDA OMNICOLL fraction collector is the possibility to collect simultaneously the streams of many chromatographic columns.

Multi-stream assembly is available for 2 to 18 or even more effluents of multiple chromatography columns. The multi-channel fraction collection configuration can also be customized for low volume recipients like 96-well plates.

In the number of simultaneous chromatographic streams the OMNICOLL is not outperformed by any other fraction collector on the market.

This multi-stream assembly is technically very simple and polyvalent. It allows an easy adaptation of various experimental needs directly by the user.

For more information about the multiple stream mounting, please do not hesitate to contact us at support@lambda-instruments.com

Figure 3.8-1 Example of taking 1x3 simultaneous streams with 96-well plates
Figure 3.8-2 Example of taking 6 simultaneous streams (1x6) into 96-well plates
Figure 3.8-3 Example of fixed 20 (2 x 10) simultaneous streams configuration
Figure 3.8-4 Example of fixed 12 (2 x 6) simultaneous streams with 250 ml bottles

 

3.9 How to increase the capacity of the fraction collector?

Since several lower parts of the collector can be coupled together, the capacity of the collector can be increased many times.

The control unit can move freely from one lower unit to the other on the gear rails. Only two rectangular coupling bars are needed to keep two units together.

  • Remove the two blind bars from the rear side of the frame of the first lower part unit and the front side of the second lower part unit. Insert the coupling bars (Art. No. 6912) on their place and fix with the four bolts.
  • Be aware that you must remove the fixed stop signal from the first frame unit and place it onto the next one. You may also use the magnetic stop contact.
  • Make sure that nothing will block the movement of the control unit during the fraction collection.

The tube capacity of the LAMBDA OMNICOLL fraction collector and sampler can be easily increased by the addition of lower support. Virtually any number of capacity extensions can be added.

Figure 3.9-1 Fraction collector with extended 30 mm tube rack capacity: 3 times (= 288 fractions)